6.4 Routine Output Calibration
Setup:
Connect the test equipment to the power source as shown in Figure 6-1. If any output parameter
can‟t be calibrated refer to the Power Source Gain Adjustment in the Non-routine Output
Calibration section of the manual.
For the AC Output Calibration either the Fluke 8506A or HP 34401A may be used. If the HP
34401A is used it must be put into the slow filter mode.
Note: For purposes of calibrating the output voltage, the 135 and 150 volt ranges
are the same. In other words, the 150 volt range can be calibrated on the 135 volt
range. The same is true for the 270 and 300 volt range.
Figure 6-1: Test Equipment Hookup for Routine Output Calibration